Top of Line Corrosion (TOL) occurs in multiphase systems when the water vapor in the gas phase condenses on the pipe walls. This occurs due to heat exchange between the pipe and the surrounding environment (river water, seawater, or cold air), and this phenomenon intensifies if the pipe is not insulated or buried at a reasonable depth.
The condensed liquid then becomes enriched with the corrosive species naturally present in the gas stream. At these locations the pH usually drops, because buffering compounds such as bicarbonate or dissolved iron are absent. Carbon dioxide (CO2) reacts with the water to form carbonic acid (H2CO3), and hydrogen sulfide (H2S) can create further serious challenges. Since TOL corrosion occurs in wet gas lines, corrosion inhibitors and other protective chemicals injected into the system, such as mono-ethylene glycol (MEG), stay at the bottom of the line and cannot protect the top of the pipe.
This test is highly recommended by NACE (now AMPP) for evaluating the effectiveness of corrosion inhibitors under top-of-line corrosion conditions.
Technical Specifications of Top of Line Corrosion Test:
- Ability to raise the temperature inside the chamber up to 80 °C with the help of a corrosion-resistant heating element.
- The TOL testing machine has 5 coupon holders: four coupon holders at two different heights relative to the test fluid surface, with the ability to simulate the thermal gradient via the cooling system at the coupon top surface, and a submersible coupon for comparison.
- The coupon positions accept standard disk or flush-disk coupons (approx. 32 mm) mounted horizontally, each with its own water-cooling jacket so the back of the coupon can be kept at room temperature.
- Transparent cylindrical body made of acrylic Plexiglas, with 150 mm outer diameter, 140 mm inner diameter and a minimum height of 150 mm.
- Completely gas and water sealed.
- Has the ability to purge gas in and above the test fluid.
- The ability to monitor the temperature of test fluid, the air above the test fluid, and cooling system temperature, using two thermocouples with digital display.
- Cooling-water circulation system with water tank, heating element and thermocouple, allowing temperature monitoring and control in the range of 25 to 80 °C.